Unit Calibration
24.
As you can see the
color temperature of
Monitor #8 is 7188K.
This is your target value
for Monitor #9
25.
Measurement Complete
Unit Calibration
26.
Click Monitor #9
27.
Always set target to
600 (cd/m2)
28.
Set color temperature
to 7000K in this case. It
is the closest value the
target of 7188.
29.
Always set Gamma
value to 2.2
30.
Click Run Calibration
and wait for crosshair to
appear on Monitor #9
Unit Calibration
31.
Attach Spyder to
Monitor #9
32.
Click “Run calibration”
process can take up to 8
minutes
33.
Click “Save result”
when done
34.
Click “Calibration off”
and close software
application.
35.
Calibration complete