Harman Multimedia OnStage UD CE Test Report
2006 011114 CE AUS/NZ
25 of 72
3.8. Device Performance Criteria for Immunity Tests
Three criteria of acceptable performance are defined by EN 55024. These are as follows
o
Criterion A - The equipment shall continue to operate as intended without operator intervention.
No degradation of performance or loss of function is allowed below a performance level specified
by the manufacturer when the equipment is used as intended. The performance level may be
replaced by a permissible loss of performance. If the minimum performance level or the permissible
performance loss is not specified by the manufacturer, then either of these may be derived from the
product description and documentation, and by what the user may reasonably expect from the
equipment if used as intended.
o
Criterion B - During the test, degradation of performance is allowed. However, no change of
operating state or stored data is allowed to persist after the test. After the test, the equipment shall
continue to operate as intended without operator intervention. The performance level may be
replaced by a permissible loss of performance. If the manufacturer does not specify the minimal
performance level (or the permissible performance loss), then either of these may be derived from
the product description and documentation, or by what the user may reasonably expect from the
equipment if used as intended.
o
Criterion C - Loss of function is allowed, provided the function is self-recoverable, or can be
restored by the operation of the controls by the user in accordance with the manufacturer’s
instructions. Functions, and/or information stored in non-volatile memory, or protected by a battery
backup, shall not be lost.
For each test method, EN 55024 specifies the appropriate criterion to be met.
Nemko USA, Inc.
11696 Sorrento Valley Road, Suite F, San Diego, CA 92121
Phone (858) 755-5525 Fax (858) 452-1810
DATE
DOCUMENT NAME
DOCUMENT #
PAGE
Janurary 5,2006
Harman Multimedia OnStage UD CE Test Report
2006 011114 CE AUS/NZ
26 of 72
3.9. Electrostatic Discharge Immunity: IEC 61000-4-2 (2001)
EN 55024 specifies the IEC 61000-4-2 Standard as the basic procedure for ESD testing. The standard
configuration as outlined in IEC 61000-4-2 (2001) is used. Tabletop devices are placed on an insulated mat
on a horizontal coupling plane. Air discharges and contact charges are made to the EUT on connectors and
conducting surfaces (as illustrated in the Test Results section of this Test Report). The discharges shall be
applied in two ways:
a) Contact Discharges to the conductive surfaces and to coupling planes:
The EUT shall be exposed to at least 200 discharges, 100 each at negative and positive polarity, at a
minimum of four test pints (a minimum of 50 discharges at each point). One of the test points shall be
subjected to at least 50 indirect discharges (contact) to the center of the front edge of the horizontal-
coupling plane. The remaining three test points shall each receive at least 50 direct contact discharges. If
no direct contact test points are available, then at least 200 indirect discharges shall be applied in the
indirect mode.
b) Air Discharge at slots and apertures, and insulating surfaces:
On those parts of the EUT where it is not possible to perform contact discharge testing, the equipment
should be investigated to identify user accessible points where breakdown may occur. This investigation
should be restricted to those areas normally handled by the user. A minimum of 10 single air discharges
of each polarity and test level shall be applied to the selected test point for each area.
For further information, please refer to the technical sections in the IEC 61000-4-2 (2001) publication in
addition to the test results section and photographs of the test set-up provided in this report.
For ESD tests, EN 55024 requires that the EUT meet at least performance Criterion B for discharges of up to
±8 kV air discharge and ±4 kV contact discharge.
Nemko USA, Inc.
11696 Sorrento Valley Road, Suite F, San Diego, CA 92121
Phone (858) 755-5525 Fax (858) 452-1810
DATE
DOCUMENT NAME
DOCUMENT #
PAGE
Janurary 5,2006
Harman Multimedia OnStage UD CE Test Report
2006 011114 CE AUS/NZ
27 of 72
Figure 6. ESD Test Setup Diagram
N
OT TO
S
CALE
CONFIGURATION LEGEND
1. Test Laboratory (6 x 7 meters)
2. Vertical Conducting Wall (3 x 3 m, grounded)
3. Ground Plane (14 square meters)
4. Ground Rod extending 3 m under ground plane
5. Non-Conducting table for ESD Simulator Control Box
6. ESD Simulator Control Box on cart
7. Electro-Static Discharge (ESD) Gun (hand held, grounded to grounding rod)
8. DC power for devices
9. Ground strap with two 470kOhm resistors
10. Grounding Strap
11. Horizontal Coupling Plane, grounded to Grounding Rod
12. Insulating Mat
13. Vertical Coupling Plane
14. EUT: Computer Speaker System and Associated System
1
2
3
4
6
7
5
8
9
11
12
10
13
14
Nemko USA, Inc.
11696 Sorrento Valley Road, Suite F, San Diego, CA 92121
Phone (858) 755-5525 Fax (858) 452-1810
DATE
DOCUMENT NAME
DOCUMENT #
PAGE
Janurary 5,2006
Harman Multimedia OnStage UD CE Test Report
2006 011114 CE AUS/NZ
28 of 72
3.10. Radio Frequency Immunity: IEC 61000-4-3 (2002)
EN 55024 specifies the IEC 61000-4-3 Standard for radio frequency (RF) immunity requirements and test
methods for equipment that is required to withstand electromagnetic (EM) fields. The RF immunity test
entails subjecting the equipment under test to a uniform field of radiated electromagnetic energy of a specified
field strength and frequency, and monitoring the functionality of the device as the frequency is swept over a
specified frequency range.
The specification limits and technical parameters for testing are outlined in the IEC 61000-4-3 (2002)
Standard. This edition of the publication specifies a transmit antenna to EUT distance of 3m and a frequency
range of 80 MHz to 1000 MHz (80% amplitude modulated at a 1 kHz rate). The standard configuration as
outlined in IEC 61000-4-3 (2002)
is
used. The EUT is set up inside a shielded, semi-anechoic chamber with a radiating antenna at a distance of 3
meters from the EUT. For further information, please refer to the technical sections in the IEC 61000-4-3
(2002)
publication in addition to the test results section and photographs of the test set-up provided in this report.
For radio frequency immunity tests, EN 55024 specifies that the EUT meet performance Criterion A for a
minimum field strength of 3 V/m.