Sharp MX-PEX1 (serv.man14) User Manual / Operation Manual ▷ View online
C
ALIBRATION
53
5 Choose the appropriate options and click Print.
Page Type:
Choose 21 or 34 Sorted Patches.
Paper Size:
Automatically set to LTR/A4 for 21 Sorted Patches. Choose 11
x
17/A3 or
12
x
18/SRA3 for 34 Sorted Patches.
Input Tray:
Specify the paper source.
6 Retrieve the measurement page from the MFP.
7 Click Measure in the Get Measurements pane.
The Measurement Options dialog box displays the specified Page Type and Paper Size
options.
options.
8 Click Measure.
Follow the on-screen instructions to feed the measurement page through the DTP32.
N
OTE
:
The DTP32 Series II does not have an edge guide for the page. Position the page so
that the strip to be measured is lined up directly under the alignment mark.
The Status field displays instructions for selecting the port and feeding the measurement page
through the DTP32 four times, once for each color strip.
through the DTP32 four times, once for each color strip.
DTP32
C
ALIBRATION
54
9 When the Status field indicates that the measurements were read successfully, click Accept.
10 Click OK in the Measure dialog box.
11 Click Apply in the Calibrator dialog box to implement the new calibration set.
12 Click OK in the Information dialog box.
This completes the calibration process.
Calibrating with the DTP41
Using the DTP41 spectrophotometer, you can measure color patches generated by the MFP
and automatically download the measurements to the Fiery X3eTY2.
and automatically download the measurements to the Fiery X3eTY2.
N
OTE
:
Before you use ColorWise Pro Tools and the DTP41 to calibrate the Fiery X3eTY2,
follow the instructions on
page 94
to calibrate the DTP41.
T
O
CALIBRATE
THE
F
IERY
X3
E
TY2
USING
THE
DTP41
1 Start Calibrator.
For instructions, see
page 46
.
2 Choose X-Rite DTP41 as the measurement method.
3 In the Check Print Settings pane, choose a calibration set.
If more than one option appears, choose the appropriate calibration set for the type of media
you use most often.
you use most often.
N
OTE
:
For this calibration to take effect, the calibration set must be associated with one or
more output profiles. The default calibration set is already associated with the default output
profile. You do not need to make any new associations.
profile. You do not need to make any new associations.
4 Click Print in the Generate Measurement Page pane.
The Print Options dialog box appears.
DTP32 Series II
C
ALIBRATION
55
5 Choose the appropriate options and click Print.
Page Type:
Choose 21 or 34 Sorted Patches.
Paper Size:
Automatically set to LTR/A4 for 21 Sorted Patches. Choose 11
x
17/A3 or
12
x
18/SRA3 for 34 Sorted Patches.
Input Tray:
Choose the paper source for printing the measurement page.
6 Retrieve the measurement page from the MFP.
7 In the Get Measurements pane, click Measure.
The Measurement Options dialog box displays the specified Page Type and Paper Size
options.
options.
8 Click Measure.
9 If necessary, choose the appropriate COM port for the DTP41 from the Available Ports menu,
and then click Start Measurements.
C
ALIBRATION
56
10 Feed the measurement page into the DTP41, starting with the cyan strip.
Align the page in the measurement page slot of the DTP41 so that the appropriate color
column is centered below the alignment mark (see the diagram on
column is centered below the alignment mark (see the diagram on
page 91
). Insert the page
about three inches, past the front idler rollers, until it rests against the rear drive rollers and
you feel resistance.
you feel resistance.
11 Click Read Strip.
The DTP41 pulls the measurement page through automatically.
After the measurement page is read, Calibrator instructs you to insert and align the page again
for the next color.
for the next color.
12 Repeat the measurement process for the magenta, yellow, and black strips.
A check mark appears on the circle of the color just read, and the Status field indicates the
color was measured successfully.
color was measured successfully.
Use the Previous and Next buttons to reverse or advance to another step in the measurement
sequence.
sequence.