Sharp MX-2630 (serv.man6) Handy Guide ▷ View online
SHARP
FIRMWARE UPDATE
MX-2630N Handy Guide NOV’17
Sharp Office Solutions Group
Page 37
/75
NETWORK CONNECTION (MFP WEB PAGE)
1. Open a web browser and enter the following URL :-
http://ipaddress/service_login.html or http://ipaddress/service_account.html
Example: http://172.16.0.126/service_login.html
Note: If User Authentication (username & password) is enabled, open the device webpage and
“login” using the service username and password (as below):-
2. Select/Enter the “Service” Username and Password to access the Service Webpage and click Login:-
Default
Username:
service
Password:
service
3. When the service webpage opens, click Update of Firmware:-
FIRMWARE UPDATE
SHARP
Page 38
/75
MX-2630N Handy Guide NOV’17
Sharp Office Solutions Group
4. Click Browse to navigate to the location of the updated firmware file and select Open:-
5. Press Execute to start the firmware update.
Note: Once the machine has entered “service download mode”, the update will commence and the
progress will be displayed.
6. When the firmware upgrade is complete, click Reboot to execute a remote restart of the machine.
Firmware Upgrade is complete
SHARP
FIRMWARE RECOVERY
MX-2630N Handy Guide NOV’17
Sharp Office Solutions Group
Page 39
/75
In the event of a power failure during a firmware update, the device may fail to “reboot” because of corrupted
firmware. Using the CN Update procedure described below, this function is used to update/recover the firmware by
transferring data to the MFP PWB, SCU PWB, PCU PWB, FAX PWB, and various options by means of a USB memory.
MENU KEY REFERENCE:
firmware. Using the CN Update procedure described below, this function is used to update/recover the firmware by
transferring data to the MFP PWB, SCU PWB, PCU PWB, FAX PWB, and various options by means of a USB memory.
MENU KEY REFERENCE:
The following five keys are used to perform firmware recovery mode. For MX-2630N series, touch the LCD
panel within the area shown below to select the desired key:-
panel within the area shown below to select the desired key:-
CN UPDATE - Firmware Recovery Procedure:-
1. Power off the MFP and remove the MFP Rear Cabinet & Screws.
2. Set the SCN-MFP CONTROL PWB UPDATE DIP-SW to ON:-
Note: If a Boot failure has occurred, replace the Boot FlashROM.
3. Install the USB Memory containing the MX-2630N firmware into the USB slot.
Note: The firmware file (*.sfu) must be saved on the main or one level lower directory of the USB
Memory.
Memory.
4. Power ON the MFP device - the UPDATE PROGRAM INIT PLEASE WAIT boot menu should appear
after approx 10 seconds, followed by VERSION CHECK CONFIG: information:-.
Version Check
Conf: 00050000
Conf: 00050000
Update Program Init
Please wait
Please wait
FIRMWARE RECOVERY
SHARP
Page 40
/75
MX-2630N Handy Guide NOV’17
Sharp Office Solutions Group
5. Select FIRM UPDATE Mode by using the [Menu] & [Back] keys. Using the [UP] and [DOWN] keys, navigate and
select the required firmware SFU.
Note:
Use the [OK] key when a directory name is selected to browse through to the sub folder(s) or
the [BACK] key to return to the parent folder(s).
Use the [OK] key when a directory name is selected to browse through to the sub folder(s) or
the [BACK] key to return to the parent folder(s).
6. Press [OK] to commence Firmware Update. – It may take approx. 1-2 minutes to read the SFU file.
After reading is completed, the firmware update (writing) will commence automatically.
Note:
The abbreviated name/location of the firmware that is being updated will appear in the upper
right corner.
The abbreviated name/location of the firmware that is being updated will appear in the upper
right corner.
7. If the firmware upgrade completes successfully, RESULT OK will be displayed:-
Result: NG = Update failed; Result: Not Update = Update is not executed.
8. Power off the MFP and return the SCN-MFP CONTROL PWB UPDATE DIP-SW to OFF (Default):-
9. Power ON the MFP, and confirm that the machine boots up normally.
10. Power OFF the MFP and refit the MFP PWB Rear Cabinet & Screws.
11. Check/confirm the firmware version using SIM 22-05.
CN UPDATE - Firmware Recovery Procedure is now complete.
Firm Update
> 0650K100.sfu
> 0650K100.sfu
Firm Update
Reading Data
Reading Data
Firm Update IcuM
Writing Data
Writing Data
Firm Update IcuM
Result : OK
Result : OK
Firm Update
From USB Memory
From USB Memory