Sharp PN-L703A (serv.man3) Service Manual ▷ View online
PN-L703A/L703B, PN-70TA3/70TB3, PN-L603A/L603B, PN-60TA3/60TB3
SERVICE MENU/VARIOUS SETTING TOOL OPERATING PROCEDURES 4- 43
21-1. OPERATING CONDITIONS FOR
THE FLICKER ADJUSTMENT TOOL
The operating conditions of the Flicker adjustment tool of the
information display are as follows.
OS
(32bit Only)
- Windows 8
- Windows Vista
- Windows 7
Communication
port
- LAN LAN(100 BASE-TX / 10BASE-TX)
■ When connected to the information display
with a LAN cable
Refer to the followings of [7. INSTRUCTIONS FOR LAUNCHER
TOOL] to connect the information display and the PC.
■ Connected to the information display with a LAN cable
■ How to prepare the network settings
■ How to set PC automatically with using DHCP
■ How to set PC manually without using DHCP
■ How to set Information Display with using DHCP
■ How to set Information Display without using DHCP
■ How to confirm the connection of PC and Information display
21-2. BOOTING PROCEDURES OF THE
FLICKER ADJUSTMENT TOOL
The operating conditions of the Flicker adjustment tool of the
information display are as follows.
Press [Flicker Adjustment] button of the service tool launcher and
the following message is displayed.
21-2.FLICKER ADJUSTMENT PROCEDURES
The operating conditions of the Flicker adjustment tool of the
information display are as follows.
Press [Flicker Adjustment] button of the service tool launcher and
the following message is displayed. Check to confirm that a jig is
installed, and press [OK] button.
The flicker adjustment procedures by use of the flicker adjustment
tool are as follows:
1) Press [Connection Setting] button, and perform the
communication setting.
2) Press [Flicker] button, and the adjustment pattern is displayed
on the monitor and [Flicker] button turns to yellow-green color.
In addition, the current adjustment value is stored in Reg
Value (VCOM value) and [UP+1], [UP+5], [DOWN-1],
[DOWN-5], and [Write] buttons are enabled.
PN-L703A/L703B, PN-70TA3/70TB3, PN-L603A/L603B, PN-60TA3/60TB3
SERVICE MENU/VARIOUS SETTING TOOL OPERATING PROCEDURES 4- 44
3) When [UP+1] or [UP+5] button is pressed, the Reg Value is
increased by 1 or 5 respectively (up to max. 127). When
[DOWN-1] or [DOWN-5] button is pressed, the Reg Value is
decreased by 1 or5 respectively (up to min. 0).
4) When [Write] button is pressed, the Reg Value is written as the
adjustment value.
5) When [Flicker] button is pressed again, the adjustment pattern
on the monitor disappears and [Flicker] button turns to gray
color. In addition, the Reg Value is deleted and [UP+1], [UP+5],
[DOWN-1], [DOWN-5], and [Write] buttons are disabled.
6) When [Exit] button is pressed, the tool is terminated. When the
tool is terminated with the adjustment pattern on the monitor
displayed, the adjustment pattern on the monitor is deleted.
(MEMO)
- When the tool is used in a LAN connection, if a certain time
passes without any operation, the connection is closed.
If, therefore, the adjustment is failed, press [Flicker] button and
connect a LAN again.
PN-L703A/L703B, PN-70TA3/70TB3, PN-L603A/L603B, PN-60TA3/60TB3
SERVICE MENU/VARIOUS SETTING TOOL OPERATING PROCEDURES 4- 45
22.RESET THE ASSIGNED COM PORT NUMBER FOR TOUCH PEN
When the different pen adaptors are connected to one PC, com port number is increased. (The com port number is increased to 256 with
one PC). When the pen adaptor is connected and the number is reaching to the upper limit. Perform this procedure.
22-1. HOW TO CONFIRM THE COM PORT NUMBE OF PEN ADAPTOR
Open the [Device Manager] and display the [Ports(COM and LPT1)] after connecting the pen adaptor to the USB port.
The number of “**” in [SHARP Touch Pen(COM**)] means the assigned port number of the Touch Panel.
22-2. HOW TO RESET THE COM PORT
Open the [Device Manager] and display the [Ports(COM and LPT1)] after connecting the pen adaptor to the USB port.
The number of “**” in [SHARP Touch Pen(COM**)] means the assigned port number of the Touch Panel.
1) Uninstall the touch pen driver.
2) Start up the special application for resetting the COM port number of the pen adaptor (CDMuninstallerGUI.exe)
by the administrative right.
3) When the CDMuninstallerGUI.exe is started up, the following window is opened.
Input “04DD” into the [Vendor ID] and input “97A7” into the [Product ID].
When the number of the COM port
exceeds 256, the monitor cannot be
operated correctly.
Perform the following reset procedure.
exceeds 256, the monitor cannot be
operated correctly.
Perform the following reset procedure.
PN-L703A/L703B, PN-70TA3/70TB3, PN-L603A/L603B, PN-60TA3/60TB3
SERVICE MENU/VARIOUS SETTING TOOL OPERATING PROCEDURES 4- 46
4) Click [Add] button.
5) Select the listed "VID_04DD PID_97A7" as shown in the following image and clich [Remove Devices].
6) Check that the “Device removed from system” or “This device could not be found on the system" message is displayed.
“Device removed from system” : Resetting operation is finished successfully.
“This device could not be found on the system" : The resetting operation has already been done, then, resetting operation is finished on the
way to the resetting.
7) Click the [OK] or the [Cancel] of CDMuninstallerGUI.exe.
8) Restart the PC.
9) Install the touch pen driver.
Click “Add”.
Select
“VID_04DDPID_97A7” and
click “Remove Devices”.