DOWNLOAD LG MC-7642W Service Manual ↓ Size: 1.77 MB | Pages: 38 in PDF or view online for FREE

Model
MC-7642W
Pages
38
Size
1.77 MB
Type
PDF
Document
Service Manual
Brand
Device
Microwave Oven
File
mc-7642w.zip
Date

LG MC-7642W Service Manual ▷ View online

5-18
(TROUBLE 7) Grill/Convection
CONDITION
CHECK
RESULT
CAUSE
REMEDY
1. Heater is not
operating.
Abnormal.
Normal.
Normal.
Heater.
Relay3, 5.
Abnormal.
Malfunction the
relay.
Malfunction the
heater.
Replace.  
Wire connection.
No continuity.
Open or loosen.
Replace or
connect tightly.  
Replace.  
2. Under or over
cooking in conv.
mode.
Abnormal.
Normal.
Heater.
Thermistor
Abnormal.
Malfunction the
thermistor.
Malfunction the
heater.
Replace.  
Replace.  
(TROUBLE 6) Oven does not cook properly when programmed for the set power level 
(Operates properly on HIGH)
CONDITION
CHECK
RESULT
CAUSE
REMEDY
Output is full when
you set lower
power level.
Abnormal.
Disconnect the
wire leads from
relay 2 and check
continuity relay 2.
(Operate the unit)
Defective P.C.B.
assembly.
Replace P.C.B.
assembly.  
NOTE: * Make sure the wore leads correct position.
* When removing the wire leads from parts be sure to grasp the connector not the wires.
* When removing the magnetron, be sure to install the magnetron gasket in the correct position and in
good condition.
MEASUREMENT WITH OUTER CASE
REMOVED
• When you replace the magnetron, measure for
microwave energy leakage before the outer case is
installed and after all necessary components are
replaced or adjusted.
Special care should be taken in measuring the
following parts. (Circled area of below Fig.)
- Around the magnetron
- The waveguide
MEASUREMENT WITH A FULLY
ASSEMBLED OVEN
• After all components, including the outer case, are fully
assembled, measure for microwave energy leakage
around the door viewing window, the exhaust opening,
and air inlet openings.
• Microwave energy leakage must not exceed the values
prescribed below.
NOTE: Leakage with the outer case removedless than
5 mW/cm.sq. Leakage for a fully assembled
oven (Before the latch switch (primary) is
interrupted) with the door in a slightly opened
position-less than 2 mW/cm.sq.
NOTES WHEN MEASURING
• Do not  exceed meter full scale deflection.
• The test probe must be removed no faster than  
1 inch/sec (2.5 cm/sec) along the shaded area,
otherwise a false reading may result.
• The test probe must be held with the grip portion of the
handle.
A false reading may result if the operator's hand is
between the handle and the probe.
• When testing near a corner of the door, keep the probe
perpendicular to the surface making sure the probe
horizontally along the oven surface, this may possibly
cause probe damage.
RECORD KEEPING AND NOTIFICATION
AFTER MEASUREMENT
• After adjustment and repair of any microwave energy
interruption or microwave energy blocking device,
record the measured values for future reference. Also
enter the information on the service invoice.
• The microwave energy leakage should not be more
than 4 mW/cm.sq. after determining that all parts are in
good condition, functioning properly and genuine
replacement parts which are listed in this manual have
been used.
• At least once a year, have the electromagnetic energy
leakage monitor checked for calibration by its
manufacturer.
WARNING : AVOID CONTACTING ANY    
HIGH VOLTAGE PARTS
5-2
• Microwave power output measurement is made with
the microwave oven supplied at its rated voltage and
operated at its maximum microwave power setting
with a load of (1000 ± 5)g potable water.
• The water is contained in a cylindrical borosilicate
glass vessel having a maximum material thickness
of 3 mm and an outside diameter of approximately
190 mm.
• The oven and the empty vessel are at ambient
temperature prior to the start of the test.
• The initial temperature (±1) of the water is (10±2)°C.
It is measured immediately before the water is
added to the vessel. After addition of the water to
the vessel, the load is immediately placed on the
center of the shelf which is in the lowest position and
the microwave power switched on.
• The time T for the temperature of the water to rise
by a value 
T of (10±2)°K is measured, where T is
the time in seconds and 
T is the temperature rise.
The initial and final water temperatures are selected
so that the maximum difference between the final
water temperature and the ambient temperature is
5°K.
• The microwave power output P in watts is calculated
from the following formula:
P =
is measured while the microwave generator is
operating at full power. Magnetron filament heat-up
time is not included.
• The water is stirred, to equalize temperature
throughout the vessel, prior to measuring the final
water temperature.
• Stirring devices and measuring instruments are
selected in order to minimize addition or removal of
heat.
WATER LOAD
TURNTABLE
MEASUREMENT OF MICROWAVE POWER OUTPUT
DISASSEMBLY AND ADJUSTMENT
A. OUTER CASE REMOVA
1) Disconnect the power supply cord from the outlet.
2) Remove the screws from the rear and along side
edges of the case.
The outer case must be moved backward to be lifted
off.
B. POWER SUPPLY CORD
1) Remove the outer case.
2) Disconnect two terminals, and remove one screw of
the earth terminal.
CAUTION: DISCHARGE THE HIGH VOLTAGE 
CAPACITOR BEFORE SERVICING
(refer to page 2-1)
C. CONTROL PANEL ASSEMBL
1) Disconnect the leadwire from the PCB SUB ASS’Y.
2) Remove the screws for the earth and securing the
control panel.
3) Lift control panel ASS’Y from the oven by the tab
unhooked.
5-3
4187 X (
T) + 0.55 X (T
- T
0
) X M
T
• T
2
: Temperature after heating
• T
0
: Temperature of bowl
• M: Weight of bowl
D. DOOR GROSS ASSEMBLY REMOVAL
1) Open the door.
2) Remove the choke cover very carefully with a    flat-
blade screwdriver.
CAUTION : Be careful not to damage door seal plate
by screwdriver.
3) Lift up and push the door.
NOTE:
1. After replacing the door, be sure to check that the
primary switch, monitor switch, and secondary switch
operate normally.
2. After replacing the door, check for microwave energy
leakage with a survey meter. Microwave energy must
be below the limit of 5 mW/cm. (with a 275 ml water
load)
3. When mounting the door assembly to the oven
assembly, be sure to adjust the door assembly parallel
to the chassis. Also adjust so the door has no play
between the inner door surface and oven frame
assembly. If the door assembly is not mounted
properly, microwaves may leak from the clearance
between the door and the oven.
E. HIGH VOLTAGE TRANSFORMER 
REMOVAL
1) Discharge the high voltage capacitor.
2) Disconnect the leadwire from magnetron, high voltage
transformer, and capacitor.
3) Remove the screw holding the high voltage
transformer to the baseplate.
F. AIR DUCT ASSEMBLY REMOVAL
1) Disconnect the leadwire from lamp and MGT
thermostat.
2) Remove the mounting screw to the magnetron.
G. MAGNETRON REMOVAL
1) Disconnect the leadwire from the high voltage
transformer and high voltage capacitor.
2) Carefully remove the mounting screws holding the
magnetron and the waveguide.
3) Remove the magnetron ASS’Y until the tube is
clear from the waveguide.
NOTE:
1. When removing the magnetron, make sure its
dome does not hit any adjacent parts, or it may be
damaged.
2. When replacing the magnetron, be sure to install
the magnetron gasket in the correct position and
be sure that the gasket is in good condition.
3. After replacing the magnetron, check for microwave
leakage with a survey meter around the
magnetron. Microwave energy must be below the
limit of 5 mW/cm
2
. (With a 275 ml. water load).
Make sure that gasket is rigidly attached to the
magnetron. To prevent microwave leakage,
tighten the mounting screws properly, making sure
there is no gap between the waveguide and the
magnetron.
5-4
Magnetron
Air Duct
Page of 38
Display

Click on the first or last page to see other MC-7642W service manuals if exist.