Harman Kardon AVR 760 (serv.man2) EMC - CB Certificate ▷ View online
Harman Kardon Inc. Page 9 of 192
AVR760 (Harman Kardon)
EMC Test Report
Reference No.: NK08E 683.EMS
7. Immunity to RF voltages (in common mode) (S2b)
7.1 Test criteria and test procedure
Operating environment;
Temperature
23 ℃
Relative Humidity
40 %
During the test of immunity to RF voltages (in common mode) the EUT was placed 0.1 M above
a metallic ground plane of dimensions 2 x 1 M.
The criterion concerning acceptable sound interference is a wanted to unwanted audio signal
ratio of ≥40 dB at a wanted signal level of 50 mW.
The requirements for receivers, multifunction equipment and video tape equipment concerning
the immunity to RF voltages in common mode are restricted to the antenna terminals and
to the frequency range from 26 MHz to 30 MHz.
The cable linking the coupling units to the EUT was short.
A band pass filter was connected to the test system and was used for the EUT AF outputs.
All unused inputs/outputs of the coupling units was terminated with the proper resistance.
The power cable of the EUT was uniformity bundled together and as short as possible.
The measurements were performed with test software TS9980 Ver.2.1
The power cable of the EUT was uniformity bundled together and as short as possible.
The measurements were performed with test software TS9980 Ver.2.1
7.2 Measurement uncertainty
Immunity to conducted currents : +/- 2 dB
7.3 Test result immunity to conducted voltage.
The equipment does comply with the requirements.
For the measurement data, see Annex C
Harman Kardon Inc. Page 10 of 192
AVR760 (Harman Kardon)
EMC Test Report
Reference No.: NK08E 683.EMS
8. Immunity to ambient electromagnetic fields (S3)
8.1 Test criteria and test procedure
Operating environment;
Temperature
22 ℃
Relative Humidity
40 %
The EUT was placed on a non-metallic support,0.1 m above the lower conductor in the center
of the strip line.
The wanted signal was fed to all input terminals respectively.
The unwanted signal is fed to a matching network of the strip line.
The criterion concerning acceptable sound interference is a wanted to unwanted audio signal
ratio of ≥40 dB at a wanted signal level of 50 mW.
The audio output power was measured with a 0.5 - 3 kHz band-pass filter.
The ground connection of the mains filter (M) was directly connected to the Jacky.
The audio analyzer was applied without a weighting filter.
The measurements were performed with test software TS9980 Ver.2.1
8.2 Measurement uncertainty
Radiated fields : +/- 2 dB
8.3 Test result immunity to conducted voltage.
The equipment does comply with the requirements.
For the measurement data, see Annex D
Harman Kardon Inc. Page 11 of 192
AVR760 (Harman Kardon)
EMC Test Report
Reference No.: NK08E 683.EMS
9. Screening effectiveness (S4)
9.1 Test criteria and test procedure
Operating environment;
Temperature
22 ℃
Relative Humidity
40 %
The equipment under test was placed on a non-metallic table of height 0.8 M.
At the side of the antenna input terminal of the equipment under test, a non-metallic table
of length 4 M was placed at the same height to provide for movement of the measuring
device, an absorbing clamp.
An RF signal generator is placed on a third table.
The equipment under test is connected to the generator, but not connected to the mains supply.
The signal from the generator is at the test frequency and un-modulated.
9.2 Measurement uncertainty
Radiated fields : +/- 2 dB
9.3 Test result immunity to conducted voltage.
The equipment does comply with the requirements.
For the measurement data, see Annex E
Harman Kardon Inc. Page 12 of 192
AVR760 (Harman Kardon)
EMC Test Report
Reference No.: NK08E 683.EMS
10. Electrostatic Discharges (EN 61000-4-2)
10.1 Test criteria and test procedure
Operating environment;
Temperature:
23 ℃
Relative Humidity:
40 %
A ground reference plane was located on the floor, and connected to earth via a low impedance connection
The return cable of the ESD generator was connected to the reference plane. In case of a floor standing
equipment, EUT was placed on the reference plane on 10cm of insulating support
In case of table-top equipment, EUT was placed on a wooden table 0.8m above the reference planed On the
floor. A horizontal coupling plane (HCP) of 1.6*0.8 m was placed on the table, and connected to the reference
plane via a cable with a 470 ㏀ resistor located in each end (0.5 mm insulating support between EUT and HCP).
In both cases a vertical coupling plane (VCP) of 0.5*0.5 m was located 10 cm from the EUT’s sides.
The VCP was connected to the reference plane in the same matter as the HCP.
The test was done by applying contact and air discharges to the EUT itself, and contact discharges to the
HCP and VCP. When applying the discharges to the HCP, the tip of the generator was located at each side of the
EUT, at 10 cm distance. When applying the discharges to the VCP the tip of the generator was located at the
middle edge of the VCP. The VCP was located 10cm from each side of the EUT.
Contact discharges of ±4 kV were applied to various points of the EUT at conductive surfaces, and to the
HCP/VCP. Air discharges of ±8 kV were applied to various points of the EUT at non-conductive surfaces.
Ungrounded equipment, or ungrounded part(s) of equipment, cannot discharge itself similarly to class I
mains-supplied equipment. If the charge is not removed before the next ESD pulse is applied, it is possible
that the EUT or part(s) of the EUT be stressed up to twice the intended test voltage.
Therefore, double-insulated equipment could be charged at an unrealistically high charge, by accumulating
Several ESD discharges on the capacitance of the class II insulation, and then discharge at the breakdown
voltage of the insulation with a much higher energy. In the case of class II & III equipment.
To avoid of unrealistically high charge, it is possible to ionize “using ionizer” before each discharge.
The return cable of the ESD generator was connected to the reference plane. In case of a floor standing
equipment, EUT was placed on the reference plane on 10cm of insulating support
In case of table-top equipment, EUT was placed on a wooden table 0.8m above the reference planed On the
floor. A horizontal coupling plane (HCP) of 1.6*0.8 m was placed on the table, and connected to the reference
plane via a cable with a 470 ㏀ resistor located in each end (0.5 mm insulating support between EUT and HCP).
In both cases a vertical coupling plane (VCP) of 0.5*0.5 m was located 10 cm from the EUT’s sides.
The VCP was connected to the reference plane in the same matter as the HCP.
The test was done by applying contact and air discharges to the EUT itself, and contact discharges to the
HCP and VCP. When applying the discharges to the HCP, the tip of the generator was located at each side of the
EUT, at 10 cm distance. When applying the discharges to the VCP the tip of the generator was located at the
middle edge of the VCP. The VCP was located 10cm from each side of the EUT.
Contact discharges of ±4 kV were applied to various points of the EUT at conductive surfaces, and to the
HCP/VCP. Air discharges of ±8 kV were applied to various points of the EUT at non-conductive surfaces.
Ungrounded equipment, or ungrounded part(s) of equipment, cannot discharge itself similarly to class I
mains-supplied equipment. If the charge is not removed before the next ESD pulse is applied, it is possible
that the EUT or part(s) of the EUT be stressed up to twice the intended test voltage.
Therefore, double-insulated equipment could be charged at an unrealistically high charge, by accumulating
Several ESD discharges on the capacitance of the class II insulation, and then discharge at the breakdown
voltage of the insulation with a much higher energy. In the case of class II & III equipment.
To avoid of unrealistically high charge, it is possible to ionize “using ionizer” before each discharge.
Instrumentation used during this session:
Instruments
Manufacturer
Model
Serial Number
Due to Calibration
ESD Simulator
Noiseken
ESS-2000
0199C02406
2008. 12
Discharge Gun
Noiseken
TCP-815P
1199C02485
2008. 12
10.2 Test Parameters
Port:
Enclosure
Test Engineer:
J. Y. Yu
Basic Standard: EN 61000-4-2:1998+A2: 2001
Performance Criteria:
B
Test Level:
Air Discharge:
±4, 8 kV
Contact Discharge:
±2, 4 kV
Click on the first or last page to see other AVR 760 (serv.man2) service manuals if exist.